Product information: This instrument uses laser and electricity; There is a risk of injury or damage to the instrument when used incorrectly. In order to protect the user and the in...
Online Message
Product Overview
Product Information
The instrument uses laser and electricity; There is a risk of injury or damage to the instrument when used incorrectly. In order to protect the user and the instrument, any unauthorized operation process must be approved by the manufacturer's maintenance engineer or application expert.
Environmental conditions:
Our designed infrared spectrometer series and accessories are used indoors at altitudes ranging from 0 to 2000 meters (or 6500 feet).
Their reliable operating temperature ranges from 16 ℃ to 25 ℃ (60F to 78F). It can be temporarily placed below -10 ℃ (e.g. during transportation) without reducing its reliability.
Infrared spectroscopy instruments need to work in a dry environment because the beam splitter, detector, and other components may be corroded by moisture. Maintain a range of non freezing moisture between 20% and 60%. Therefore, infrared spectrometers need to be sealed and the indoor air dried. If you have any further questions about the environmental conditions of the equipment, please contact our customer service department
The door.
System configuration requirements:
The CPU frequency is above 1GHz;
● Memory above 1GB;
The computer is required to have a U 2.0 interface.
The hard disk space is above 1 GB;
● Video memory above 4M;
● Supports VGA mode display with a resolution of 1024 × 768 or higher;
Mouse, keyboard, and CD-ROM;
● Printer;
Windows XP and Windows 7 operating systems. If it is a Windows 98 operating system, it must have Internet Explorer 5.0 or higher.
Main technical parameters:
1. Wave number range 7800cm ⁻¹ -350cm ⁻¹
2. Resolution: 1.0cm ⁻¹
3. Transmittance repeatability: 0.5%
4. Light source: High intensity silicon carbide light source
5. Beam splitter: multi-layer coated potassium bromide
6. Monitor: High sensitivity DTGS
7. Dimensions: 500 × 420 × 210 (mm)
8. Weight: 24kg
Note: This product complies with the standard GB/T 21186-2007 Fourier Transform Infrared Spectrometer.